• English
  • Lithuanian
excl VAT

SNAIGĖ LF10SD-P6000E2 laboratory freezer

Product code: LF10SD-P6000E2XXXXXZ1LT4A

Available quantity 2 weeks

100 / 98 l, −25…−10 °C, temperature logging and Bluetooth. Static cooling, fixed shelves and manual defrost.
€1,52339 inc VAT
€1,25900 excl VAT
Have any questions about this item? Ask
Manufacturer: Snaige

SNAIGĖ LF10SD-P6000E2 laboratory freezer

A laboratory freezer for temperature-sensitive samples and reagents. The adjustable range is −25…−10 °C and gross / net volume is 100 / 98 l. Check the storage requirements of your materials before selection.

Static cooling and fixed shelves

The freezer uses static cooling. The evaporator shelves have fixed positions and cannot be height-adjusted. Frost accumulates over time, so the cabinet must be defrosted manually at intervals. Arrange suitable temporary storage for its contents during defrosting.

Monitoring and alarms

The electronic controller records temperature data for review and transfer through Bluetooth and the CONTROLLA app. Visual and audible alarms indicate temperature deviations and power failure. Backup power supports the controller, recording and alarms for at least 24 hours during a power outage. It does not power the refrigeration system.

The cabinet has a probe access port, external alarm contact and lockable door. The static cooling system requires manual defrost.

The GSM remote alarm system is ordered separately. Please confirm compatibility and pricing before ordering.

Manufacturer specifications and manuals

Capacity and internal fittings
Gross / Net Volume, l 100 / 98
Shelves No & Type 3 fixed evaporator shelves
Controls and functions
Controller Type Electronic
Temperature alarm Audible and visual
Volt-free alarm contact Yes
Dimensions and weight
Depth, mm 620
External Dimensions (W×D×H), mm 560 × 620 × 870
Height, mm 870
Width, mm 560
Energy and electrical data
Voltage / Frequency, V/Hz 220–240 V / 50 Hz
General information
Equipment type Laboratory freezer
Manufacturer SKU – standard model LF10SD-P6000E2XXXXXZ1LT4A
Medical safety and monitoring functions
Temperature data logger Integrated
Data communication interfaces Bluetooth, RS485
Power failure alarm Yes
Controller and alarm backup power and duration, h At least 24 hours
Temperature logging during power failure Yes
Refrigeration and ventilation system
Cooling Type Static
Defrost Type Manual
Refrigerant R600a
Temperature and performance
Temperature Range, °C −25 to −10