This group covers frozen storage of samples, reagents and other materials subject to their storage requirements. The current catalogue includes four TEFCOLD SE chest freezers (−45 to −5 °C). Match the entire operating range, usable capacity and monitoring needs.
The range is being expanded with three upright BOLARUS SLMB models. Compact SLMB 100 operates from −40 to −30 °C and full-height SLMB 500 and 700 from −45 to −30 °C; listed capacities are 100, 500 and 700 l, with net volume to be confirmed. SLMB 500/700 feature four fixed shelves and an AISI 304 interior. These cabinets help separate sample groups on shelves where front access is important. TEFCOLD SE chest freezers remain an option where a wider −45 to −5 °C operating range or top loading is required.
SLMB availability and service arrangements are still being confirmed. Before ordering, verify BOL50 refrigerant service support, the exact defrost method and any additional logging configuration. SLMB 100 does not cover −45 °C; select a model whose entire operating range meets the requirement.
|
Model |
Net volume |
External W×D×H |
Energy use |
Typical selection logic |
|---|---|---|---|---|
|
SE10-45 |
152 l |
734 × 705 × 945 mm |
3.45 kWh/24 h |
Most compact option where floor space is limited |
|
SE20-45 |
230 l |
984 × 705 × 945 mm |
3.55 kWh/24 h |
Medium storage capacity |
|
SE30-45 |
323 l |
1284 × 705 × 945 mm |
5.15 kWh/24 h |
Higher capacity for larger stored quantities |
|
SE40-45 |
392 l |
1504 × 705 × 945 mm |
5.52 kWh/24 h |
Largest active SE model in this category |
All four active SE models carry the same stated temperature range of −45 to −5 °C. Within this series, model choice is therefore mainly about capacity and physical size, while the setpoint must follow the requirements of the material being stored.
If your process requires temperatures below −45 °C, the SE range is not suitable. A temperature setting alone also does not prove compliance with a specific medical or laboratory standard.
SE10-45 provides 155 / 152 l gross / net volume and measures 734 × 705 × 945 mm. TEFCOLD lists 3.45 kWh/24 h energy use and 250 W input power.
It is the logical choice where the stored quantity is small and floor space is limited.
SE20-45 provides 235 / 230 l gross / net volume, measures 984 × 705 × 945 mm, uses 3.55 kWh/24 h and has 300 W input power.
It is the mid-size step between the compact SE10-45 and the higher-capacity SE30-45 / SE40-45.
SE30-45 provides 330 / 323 l gross / net volume, measures 1284 × 705 × 945 mm, uses 5.15 kWh/24 h and has 500 W input power.
Choose it when substantially more storage volume is needed but the 1.5 m wide SE40-45 would be unnecessarily large.
SE40-45 is the largest active SE model in the category: 400 / 392 l gross / net volume, 1504 × 705 × 945 mm, 5.52 kWh/24 h energy use and 600 W input power.
Check not only the floor footprint but also lid-opening space and comfortable access across the full 1.5 m cabinet width.
TEFCOLD specifies an electronic thermostat with alarm on all four models. This allows the operating temperature to be set electronically and provides the controller’s alarm function.
Do not infer functions that are not stated on the current product pages, such as cloud monitoring, GSM alarms or integrated continuous data logging.
The SE range uses static cooling, with no internal fan-driven air circulation listed. Loading pattern and container placement can therefore influence temperature distribution inside the chest.
If your process requires documented temperature-uniformity performance throughout the entire storage volume, check the relevant manufacturer or validation documentation rather than relying only on the overall temperature range.
All four models use manual defrost. This keeps the refrigeration design simple but requires planned defrosting and an internal procedure for protecting or transferring stored materials during the defrost period.
Defrost frequency will depend on ambient humidity, lid-opening frequency and operating conditions.
Each SE model has one hinged solid lid and a lock. The solid construction avoids the additional heat gain of a glass display lid, while the lock provides basic physical access control.
The mechanical lock is not the same as audited electronic access control; where traceability is required, a separate solution may be necessary.
TEFCOLD specifies a powder-coated steel interior, one white wire basket and four castors, two with brakes. The basket can keep frequently accessed smaller items separate from the main storage volume.
Plan the actual container or box sizes before purchase because the ergonomics of a horizontal chest are very different from an upright laboratory cabinet.
TEFCOLD lists climate class 3 for all four models. This should be considered together with the installation-room conditions and the instructions for the individual model.
Avoid strong external heat sources and leave the ventilation / service clearances required by the manufacturer.
SE10-45, SE20-45, SE30-45 and SE40-45 use R290 refrigerant.
R290 is flammable, so refrigeration-system service or leak-related work should be performed by qualified personnel in accordance with the manufacturer instructions and applicable safety requirements.
Select equipment according to the actual material, temperature and monitoring requirements. A temperature range alone does not establish suitability for every medicine, vaccine or laboratory sample. Where required, confirm the selected model’s temperature uniformity, logging and validation documentation before purchase.
The SE range has a clearly defined feature set: −45 to −5 °C, electronic thermostat with alarm, static cooling and manual defrost. It is not a universal solution for every medical application.
If the project requires lower temperatures, continuous data logging, remote alarm, validated temperature uniformity, specific certification or integrated backup systems, verify those requirements separately before selecting an SE model.
All four active models are specified at −45 to −5 °C.
Main differences are net volume, cabinet width, input power and energy use. The general temperature and feature concept remains the same.
SE10-45: 152 l net and 734 mm wide.
SE40-45: 392 l net and 1504 mm wide.
TEFCOLD specifies an electronic thermostat with alarm on all four models.
Yes. The current product specifications list a lock and one hinged solid lid.
No. TEFCOLD specifies static cooling.
No. The SE series uses manual defrost.
Yes. Each model is specified with one white wire basket.
They have four castors, two with brakes.
All four active SE models use R290.
No. Suitability depends on the exact storage, monitoring and validation requirements of the material and the documentation for the selected model.
No. Both have the same −45 to −5 °C range; SE40-45 mainly provides more storage volume.